MPT-200C - 200W +5V & +/-15V Medical Open Frame Power Supply
Features
Input Voltage
90~264 VAC, 127~370VDC
Input Frequency
47~440Hz
Power Factor
PF>0.95/230VAC PF>0.98/115VAC, full load
AC Input Current
3.5A at 115VAC, 60Hz or 1.6A at 230VAC, 60Hz full load
Inrush Current
<25a 115vac="115vac" 230vac="230vac" at="at" cold="cold" or="or" p="p" start="start">Setup, Rise Time
1000ms, 20ms / 230VAC at full load , 3000ms, 20ms / 115VAC at full load
Hold-Up Time
16ms / 230VAC at full load, 16ms / 115VAC at full load
Leakage Current
<180 264vac="264vac" at="at" earth="earth" for="for" leakage="leakage" p="p"><100 264vac="264vac" at="at" for="for" leakage="leakage" p="p" patient="patient">DC Output ADJ Range
-5% to +10% rated output, on first output
Over Load Protection
120~160% rated load
Auto recovery once fault condition removed
Over Voltage Protection
115%~135% rated load, on first output
Over Temperature Protection
At 95°C. Will shut down and recover automatically after temperature goes down
Power Good/Fail Signal
TTL logic high for power good and TTL low for power fail
Remote Control
RC+/RC-:0 ~ 0.8V=power on; 4 ~ 10V=power off sink current<4 ma="ma" p="p">Working Temperature
-20°C to 70°C (refer to derating curve)
Working Humidity
20~90% RH non-condensing
Storage Temperature
-40°C to +85°C
Storage Humidity
10~90% RH
Temperature Coefficient
±0.04%/°C (0~50°C) full load
Vibration
2G 10Hz ~500Hz 10 min cycle
Medical Safety Standards
UL60601-1, TUV EN60601-1 approved
Commercial Safety Standards
Complies with UL60950-1, TUV EN60950-1
Withstand Voltage
I/P-O/P: 4000VAC I/P-FG, O/P-FG: 1500VAC
Isolation Resistance
I/P-O/P, I/P-FG, O/P-FG : >100M Ohms
EMI
Compliance to EN55011 Class B, EN55022 Class B, EN61000-3-2, EN61000-3-3
EMS
Compliance to EN61000-4-2 Level 3, 8KV, EN61000-4-2 Level 2, 4KV
EN61000-4-3 Level 2, 3V/m , EN61000-4-3 Level 3, 10V/m
EN61000-4-4 Level 2, 1KV/5KHz, EN61000-4-4 Level 3, 2KV/5KHz
EN61000-4-5 Level 4, 2KV/Line-Line, 4KV/Line-Earth
EN61000-4-6 Level 2, 3Vrms/m, Level 3, 10Vrms/m
EN61000-4-8 Level 2, 3A/m, EN61000-4-8 Level 3, 10A/m, EN61000-4-11 Compliance
ENV50204 Level 2, 3V/m, 900MHz, ENV50204 Level 3, 10A/m, 900MHz
MTBF
262,100 hours min MIL-HDBK-217F
Dimensions
177.8 x 107.2 x 35.5mm
MPT-200C output Specifications
Voltage (max/rated)
Current (amps)
Power (watts)
R&N (mV)
Efficiency (%)
5V
20A
200W
80mV
80%
15V
5A
150mV
-15V
2A
80mV
4>100>180>25a>